Rejoining Separated fluid in Thermometers

If you occupational in a materials testing lab, you’ve most likely experienced fail of a liquid-in-glass (LiG) lab thermometer because of a separated shaft of mercury or indicating fluid. Separation can happen throughout shipping or from improper storage or use. If the liquid in the shaft is separated, all succeeding readings will certainly be inaccurate. This typical occurrence is not usually a reason to return or discard the thermometer. Many of the time, the separated liquid column deserve to be rejoined by cautiously following specific steps, allowing the thermometer to be went back to service.

You are watching: How to reset a mercury thermometer


*

Photo Credit: AASHTOresource.org


This blog post will explore usual practices arisen by others to reunite a separated column in bothmercury-in-glass thermometers and also spirit-filled thermometers.

Practicing proper vertical warehouse of LiG thermometers in athermometer warehouse rack minimizes column separation issues, stays clear of breakage, and keeps the tools organized and out the the way.

6 Do"s and Don"ts for Separated Mercury-Filled Thermometers

*

#1: Attempting come rejoin a separated tower on her own

Do be experienced and also careful if you attempt this process. Rejoining a be separated mercury pillar is not an “I’ll offer it a try” endeavor. Mercury is a hazardous material and breakage that a glass thermometer is thought about a mercury spill requiring strict measures to prevent person exposure. It requires expertise and also skill, keeping safety a priority.Don"t hesitate to ask because that assistance. If you have actually not done this procedure before, contact the manufacturer or a calibration lab experienced in the procedure for guidance and also help. Inspect out the links in the More Information section at the finish of this blog.

#2: Remember: the mercury is a hazardous material!

Do exercise excessive care and also caution as soon as rejoining a separated column in a mercury thermometer.

#3: remember the regulations and also safety protocols

Do monitor regulatory firm and internal laboratory policies and also safety procedures totally to prevent breakage and avoid a hazardous mercury spill. If a spill occurs, inform your safety and security officer immediately.Don"t attempt these measures on a mercury thermometer without having a complete mercury spill kit top top hand.

#4: Cooling is the preferred method

Except for deep immersion thermometers, this technique can be supplied with many mercury thermometers to reunite a separated column.

Do shot the cooling an approach first. that is considered the safest means to rejoin the be separate mercury, and also can it is in performed utilizing the adhering to easy-to-find items:Solid Carbon Dioxide (CO2), also known as dry iceAlcohol or AcetoneGloves and Safety GogglesDon"t touch the bulb once the cooling an approach has to be completed until it has warmed sufficient to allow the mercury to circulation from the bulb right into the capillary.

The objective of this procedure is to rejoin the separations and ensure that the gas remains above the mercury. Holding the thermometer upright, slowly and gradually immerse just the pear of the instrument into a equipment of dry ice and alcohol consisted of in a small lab beaker or one more such upright container. Throughout this gradual insertion, the mercury should start to solidify. Be cautious to not immerse the stem or shaft of the thermometer into the cooling equipment or allow the mercury come freeze together that can cause the bulb to fracture. If necessary, relocating the pear in and also out that the solution many times deserve to slow down and much better control the cooling action. When all the be separated portions have been lugged into the bulb, take it the thermometer out of the container and swing it in a short arc to pressure the mercury right into the bulb.

Placing the thermometer upright in one upright thermometer warehouse rack as it return to room temperature keeps the nitrogen gas on optimal of the mercury. Save thermometers properly when not in usage reduces the likelihood of the mercury pillar separating again.

#5: utilizing the Heating method for rejoining

Do usage this method only because that thermometers with a maximum range of 260°C (500°F), and also with expansion chambers1 able come accommodate the separation and also a part of the key column.1The expansion chamber is at the top of the capillary in a LiG thermometer. The chamber avoids the buildup of pressure if the best temperature is exceeded and the fluid rises previous the height of the scale.Don"t ever heat the bulb v an open up flame or to fill the growth chamber more than two-thirds full, or use the heat technique to rejoin columns in thermometers through a selection in excess of 260°C (500°F). Any type of of these can damage the thermometer or reason the pear to burst.

In enhancement to safety gloves, goggles, and also other vital safety gear, a large beaker is v a fluid heated below its flashpoint, but never over 260°C (500°F). Immerse the bulb and stem into the liquid and stir v the thermometer till the separation and component of the main column just start to get in the expansion chamber. The growth chamber should never be much more than two-thirds full. As soon as the liquid is in the chamber, eliminate the thermometer native the beaker and also hold that upright vertically. Tenderness tap it into the palm of her gloved hand come reunite the column. Enable the thermometer come cool slowly. Again, keeping unused thermometers in a warehouse rack will aid prevent future separation.

Note:An alternate to the maker is to use a well-stirred, temperature-regulated fluid bath.

#6: Gas balloon in thermometers v a convulsion chamber

It’s precious noting that mercury thermometers through a convulsion chamber2 deserve to be susceptible to separations, as well as bubbles the gas. Rough handling during shipping or regular jarring the the tool in the rap can cause gas bubbles in the bulb.2Some LiG thermometers with scale ranges that begin over ambient temperatures have actually a contraction chamber situated just listed below the key scale that the device. This enlargement in the capillary prevents excess fluid from start the bulb at reduced temperatures and shortens the distance the fluid must travel when measuring higher temperatures.

Do research the guideline of the bulb routinely for any gas bubbles. Don"t madness the thermometer pear on a difficult surface. Doing for this reason can cause the gas to rest up into many smaller bubbles and also worsen the situation.

To resolve the issue, use the cooling an approach described over to cool a part of the bulb till a large bubble that gas beginning it. Remove the thermometer indigenous the cooling container, and also hold the upright with the bulb side up, and also begin to tenderness tap it against the palm of her hand. Together gas is lighter than mercury, the gas bubble will climb up the size of the bulb. Invert and also tilt the thermometer so it now stems up and also roll it around to allow the gas balloon to come into contact with the bulb surface. The huge bubble will certainly pick up the other smaller gas particles together they come in contact. Proceed rolling the thermometer in this manner until all smaller gas particles have actually been picked up. When all the bubbles room absorbed, continue to monitor the Cooling method described above to get the gas positioned over the mercury in the thermometer. Be mindful to permit the bulb heat to the suggest that the mercury is no longer solid prior to touching it.

Finally, keep in mind the there are federal, state, and local ordinances that manage mercury instruments. Regulations are constantly changing and companies might not be enabled to delivery mercury thermometers to some areas.

Tips for Separated Non-Mercury filled Thermometers

*

Fluid obelisk separations can occur just as quickly in non-mercury or spirit-filled LiG thermometers. Back rejoining separations in these tools don’t attitude as good a risk, the procedure still requires experience and expertise.

Tip #1:A simple and safe technique is using a centrifuge to attract the liquid under the capillary. The centrifuge cup have to be deep enough so the centrifugal pressure is listed below the liquid column. If the centrifugal pressure is not listed below the column, the column will split and only pressure a portion of the fluid down, with the rest migrating up into the growth chamber. Noodle padding must be put in the bottom the the cup come prevent damage to the bulb. Ar the thermometer bulb down in the cup and turn top top the centrifuge to conveniently force every the liquidthrough the separation – this need to take simply a few seconds.Tip #2: If no centrifuge is available, one more option is to host the thermometer upright and gently tap the stem the the thermometer against the palm of your hand. As you space tapping, save watching the liquid above the separation – it must break far from the capillary wall surface and run down to join the key column. Proceed the tapping activity and observation until all the liquid has rejoined the key column. If fluid did do its method into the growth chamber, continue this process until every the fluid is reunited.